bio-flow-cytometry-cytometry-qc

Comprehensive quality control for flow cytometry and CyTOF data. Covers flow rate stability, signal drift, margin events, dead cell exclusion, and batch QC. Use when assessing acquisition quality or identifying problematic samples before analysis.

personAuthor: jakexiaohubgithub

Version Compatibility

Reference examples tested with: flowAI 1.32+, PeacoQC 1.12+, flowCore 2.14+, flowDensity 1.36+, CATALYST 1.26+.

Before using code patterns, verify installed versions match. If versions differ:

  • R: packageVersion('<pkg>') then ?function_name to verify parameters

Counterintuitive defaults to confirm: flowAI checks are FR/FS/FM (FM = dynamic range, not "flow"); PeacoQC MAD/IT_limit are LESS strict when HIGHER. Verify with ?flow_auto_qc and ?PeacoQC before tuning.

Cytometry QC

"Run quality control on my cytometry data" -> Detect and remove acquisition artifacts (flow-rate instability, signal drift, margin events, dead cells, CyTOF doublets) on the Time axis, then flag outlier samples.

  • R (flow/spectral): flowAI::flow_auto_qc(), PeacoQC::PeacoQC() (+ RemoveMargins())
  • R (mass): CATALYST::normCytof() beads + Gaussian/DNA/event-length gating

The Single Most Important Modern Insight -- The Time Parameter Is the Master QC Axis, and Order Matters

Nearly every acquisition artifact - clogs, bubbles, flow-rate surges, electronics warm-up, CyTOF sensitivity decay, oxide buildup - manifests as a CHANGE IN SIGNAL versus the Time channel. flowAI, flowCut, flowClean, and PeacoQC are all, at heart, Time-vs-signal anomaly detectors; a missing or mis-scaled $TIMESTEP silently degrades or breaks all of them. Just as important is the ORDER: compensation/unmixing -> transform -> margin removal -> time-based QC -> debris/doublet/dead-cell gating -> batch normalization. Margin (boundary) events piled at a detector min/max form spurious high-density ridges that fool density-based cleaning and density gates, so they must be stripped BEFORE any density step; and time-based QC on untransformed data misbehaves because the density structure the algorithms rely on lives on the transformed scale.

Cleaning-Tool Taxonomy

| Tool | Citation | Mechanism | When to use / caveat | |------|----------|-----------|----------------------| | flowAI | Monaco 2016 Bioinformatics 32:2473 | 3 checks: flow rate (FR), signal acquisition (FS), dynamic range (FM) | classic; known AGGRESSIVE - can remove normal data | | PeacoQC | Emmaneel 2022 Cytometry A 101:325 | per-channel density peaks + MAD + isolation tree | only tool validated across flow + mass + spectral; QC engine of CytoPipeline | | flowCut | Meskas 2023 Cytometry A 103:71 | segments Time, removes low-density/deviant segments | less aggressive than flowAI; flags whole files | | flowClean | Fletez-Brant 2016 Cytometry A 89:461 | tracks subset frequency in centered-log-ratio space | floor ~30,000 events; writes a "GoodVsBad" parameter to gate on |

Run flowAI (with the correct API)

Goal: Auto-clean a sample for flow-rate, signal-acquisition, and dynamic-range anomalies.

Approach: flow_auto_qc() returns a flowFrame of high-quality events when output=1; FM is the dynamic-range check; supply timeCh for concatenated/clock-reset files. flowAI is the time-based QC step - run it after compensation/transform/margin removal (per the ordering above), not on a raw uncompensated frame.

library(flowAI)

ff_clean <- flow_auto_qc(ff,
                         remove_from = 'all',          # FR + FS + FM
                         output = 1,                    # 1 = HQ events only; 2 = add QC param; 3 = bad-event IDs
                         ChExcludeFS = c('FSC', 'SSC'), # scatter excluded from the signal check
                         second_fractionFR = 0.1,
                         folder_results = 'qc_output')
cat('kept', nrow(ff_clean), 'of', nrow(ff), 'events\n')

Margins First, Then PeacoQC

Goal: Remove boundary events, then clean unstable time/peak structure across all channels.

Approach: RemoveMargins() strips detector-min/max events; then PeacoQC() - remember higher MAD/IT_limit = LESS strict.

library(PeacoQC)

ff_nm <- RemoveMargins(ff, channels = c('FSC-A', 'SSC-A'))   # do this BEFORE density QC
res <- PeacoQC(ff_nm, channels = marker_channels,
               MAD = 6, IT_limit = 0.55,                     # defaults; higher = less strict (counterintuitive)
               save_fcs = FALSE, plot = TRUE)
ff_clean <- res$FinalFF

Dead-Cell, Drift, and CyTOF Checks

Goal: Exclude dead cells, detect per-channel drift, and apply CyTOF-specific gates.

Approach: Viability dye threshold (bimodal); per-time-bin median slope for drift; for CyTOF use DNA intercalator + Gaussian/event-length; EQ-bead-median-vs-Time is the primary CyTOF drift readout (see bead-normalization).

expr <- exprs(ff)
# dead cells take up more viability dye -> cut at the bimodal density VALLEY (data-driven), not a fixed quantile
dead_cut <- flowDensity::deGate(ff, channel = 'Live_Dead')
live <- expr[, 'Live_Dead'] < dead_cut

# CyTOF single-cell gates
if ('Event_length' %in% colnames(expr)) {
    keep <- expr[, 'Event_length'] >= 10 & expr[, 'Event_length'] <= 75   # confirm range per instrument
}
dna <- grep('Ir191|Ir193', colnames(expr), value = TRUE)             # intercalator-positive = nucleated

Calibration and Standardization (cross-study comparability)

A discovery analyst often skips this, but cross-experiment/cross-site MFI comparison is meaningless without it (Maecker & Trotter 2006 Cytometry A 69:1037):

  • MESF / MEF / ERF beads express intensity in molecules-of-equivalent-fluorochrome - comparable across instruments and time (NIST/ISAC standard; PE/Pacific Blue use ERF surrogates).
  • Quantibrite PE (defined PE molecules/bead, ~1:1 conjugation) converts MFI to antibodies-bound-per-cell / receptor density (bead values are LOT-dependent).
  • CS&T / 8-peak rainbow beads for daily QC (laser delay, area scaling, linearity).
  • Peak-2 / voltration: run a dim particle across PMT voltages, pick the CV-vs-voltage inflection = minimum voltage for optimal resolution. This is why MIFlowCyt mandates reporting voltages.

Batch-Level Outlier Flagging

Goal: Flag samples whose event count, flow stability, or marker medians deviate from the batch.

Approach: Per-file metrics + MAD-based bounds; track an anchor/reference sample if present.

qc <- do.call(rbind, lapply(fcs_files, function(f) {
  ff <- read.FCS(f); e <- exprs(ff)
  data.frame(file = basename(f), events = nrow(ff),
             med_signal = median(apply(e, 2, median)))
}))
qc$outlier <- abs(qc$events - median(qc$events)) > 3 * mad(qc$events)

Per-Method Failure Modes

Density QC on un-margin-removed data

Trigger: PeacoQC/flowClean before RemoveMargins. Mechanism: axis pile-ups are false high-density ridges. Symptom: real events removed near the boundary, or margins kept. Fix: remove margins first.

flowAI over-removal

Trigger: default flowAI on a low-rate or short acquisition. Mechanism: FR check flags normal slow segments. Symptom: large unexplained event loss. Fix: raise second_fractionFR; inspect the HTML report; consider flowCut/PeacoQC.

QC on untransformed/uncompensated data

Trigger: running QC on raw linear values. Mechanism: high-intensity tail dominates density. Symptom: misplaced anomaly calls. Fix: compensate + transform first.

Time axis missing/reset

Trigger: concatenated files, some sorters. Mechanism: no usable Time. Symptom: flow-rate check fails or is meaningless. Fix: timeCh= or reconstruct; otherwise skip time-based checks.

Quantitative Thresholds

| Threshold | Source | Rationale | |-----------|--------|-----------| | flowClean floor ~30,000 events | Fletez-Brant 2016 Cytometry A 89:461 | below this the CLR frequency tracking under-detects | | PeacoQC MAD=6, IT_limit=0.55 | Emmaneel 2022 Cytometry A 101:325 | defaults; HIGHER = less strict | | dead cells > ~10-30% | community | sample-handling flag, not a hard cutoff - report, don't auto-exclude the sample | | CyTOF retune ~ daily / per long run | instrument practice (flagged) | sensitivity decays from cone fouling/plasma drift |

Common Errors

| Error / symptom | Cause | Solution | |-----------------|-------|----------| | flow_auto_qc returns unexpected object | assuming $fcs/report list | output=1 returns a flowFrame; set output explicitly | | margins not removed by PeacoQC | expecting it built-in | call RemoveMargins() separately, first | | tuning MAD up removes more | sign confusion | higher MAD/IT_limit = LESS strict | | flowClean output unchanged | it appends a parameter | gate on the "GoodVsBad" column |

References

  • Monaco 2016 Bioinformatics 32(16):2473-2480 — flowAI.
  • Emmaneel 2022 Cytometry A 101(4):325-338 — PeacoQC.
  • Meskas 2023 Cytometry A 103(1):71-81 — flowCut.
  • Fletez-Brant 2016 Cytometry A 89(5):461-471 — flowClean.
  • Fienberg 2012 Cytometry A 81(6):467-475 — cisplatin viability reagent (CyTOF live/dead).
  • Maecker & Trotter 2006 Cytometry A 69(9):1037-1042 — controls, instrument setup, peak-2.
  • Lee 2008 Cytometry A 73(10):926-930 — MIFlowCyt reporting (voltages, clones, config).

Related Skills

  • compensation-transformation - Compensate/transform before time-based QC
  • doublet-detection - Singlet discrimination after QC
  • bead-normalization - EQ-bead drift correction for CyTOF (QC's normalization arm)
  • clustering-phenotyping - Cluster only QC-passed events
  • experimental-design/batch-design - Anchor/reference-sample design for batch QC